방사선방어용앞치마(Apron)의 차폐율과 균일성 측정을 통한 성능평가 연구

A study on performance evaluation of Apron by shielding rate and uniformity

유세종(건양대학교); 임창선(건양대학교); 심규란(건양대학교)

17권 1호, 103~109쪽

초록

In this study, we quantitatively analyzed the data by measuring the radiation shielding rate and uniformity in order to evaluate the performance of an Apron. In addition, storage conditions were also evaluated. The uniformity measurement was performed by evaluating the Apron DICOM images using a PACS program. The experiment was intended for 51 Aprons being used in three hospitals in the Daejeon area. The radiation shielding rate and uniformity were measured per lead equivalent for 0.25 mmPb, 0.35 mmPb, and 0.5 mmPb. As a result, the higher lead equivalents were, the greater differences in the non-uniformity between the top part and the bottom part became (p=0.020). In all hospitals, regarding the non-uniformity of four places in Aprons, all showed statistically significant differences (p<0.01). The average value of the transmitted radiation dose showed less difference (p = 0.005) in the bottom right than in the upper right but was statistically significant. There have been no marks of manufacturing date or the date of purchase in the Apron.

Abstract

In this study, we quantitatively analyzed the data by measuring the radiation shielding rate and uniformity in order to evaluate the performance of an Apron. In addition, storage conditions were also evaluated. The uniformity measurement was performed by evaluating the Apron DICOM images using a PACS program. The experiment was intended for 51 Aprons being used in three hospitals in the Daejeon area. The radiation shielding rate and uniformity were measured per lead equivalent for 0.25 mmPb, 0.35 mmPb, and 0.5 mmPb. As a result, the higher lead equivalents were, the greater differences in the non-uniformity between the top part and the bottom part became (p=0.020). In all hospitals, regarding the non-uniformity of four places in Aprons, all showed statistically significant differences (p<0.01). The average value of the transmitted radiation dose showed less difference (p = 0.005) in the bottom right than in the upper right but was statistically significant. There have been no marks of manufacturing date or the date of purchase in the Apron.

발행기관:
대한안전경영과학회
DOI:
http://dx.doi.org/10.12812/ksms.2015.17.1.103
분류:
안전공학

AI 법률 상담

이 논문의 주제에 대해 더 알고 싶으신가요?

460만+ 법률 자료에서 관련 판례·법령·해석례를 찾아 답변합니다

AI 상담 시작