첨가법에 의해 제작된 단관의 내면 적합도 분석
Analysis of internal fitness of single crown fabricated by additive method
김원수(대전보건대학교)
37권 4호, 229~234쪽
초록
Purpose: The purpose of this study was to analysis the internal fitness of single crown fabricated by additive method technique. Methods: Study models were fabricated. Ten crowns were manufactured by additive method technique (AM group) and another ten crowns using conventional method were manufactured (CM group). The internal fitness of specimens were measured using silicone replica technique. Silicone replicas were sectioned one times (direction: from mesial to distal). The internal fitness were evaluated using by digital microscope. Statistical analyses were performed with independent samples t-test. (α=0.05). Results: Mean(SD) of all internal fitness were 92.1(20.0)㎛ for AM group and 69.7(12.3)㎛ for CM group. Two groups were statistically significant differences (p<0.001). Conclusion: All internal gaps of AM group were ranged within the clinical recommendation (120㎛).
Abstract
Purpose: The purpose of this study was to analysis the internal fitness of single crown fabricated by additive method technique. Methods: Study models were fabricated. Ten crowns were manufactured by additive method technique (AM group) and another ten crowns using conventional method were manufactured (CM group). The internal fitness of specimens were measured using silicone replica technique. Silicone replicas were sectioned one times (direction: from mesial to distal). The internal fitness were evaluated using by digital microscope. Statistical analyses were performed with independent samples t-test. (α=0.05). Results: Mean(SD) of all internal fitness were 92.1(20.0)㎛ for AM group and 69.7(12.3)㎛ for CM group. Two groups were statistically significant differences (p<0.001). Conclusion: All internal gaps of AM group were ranged within the clinical recommendation (120㎛).
- 발행기관:
- 대한치과기공학회
- 분류:
- 기타치의학